Differential Interference Contrast (DIC)

General remarks

  • DIC allows observation of unstained specimens by generation of image contrast
  • Structures of interest must have different refractive indices (r.i.).

Beam path for DIC microscopy

Keep in mind: there are dedicated DIC prism / objective combinations. Check appropriate combinations from the Objective list of the manual or the Facility web site.

1. Upright microscope in laser scanning mode (from top to bottom)

  • Light source (laser)
  • No Polarizer (laser has to provide correctly polarized light)
  • DIC prism at objective
  • Objective
  • Sample
  • Condenser front lens
  • DIC prism (condenser)
  • Analyzer (polarizing filter)
  • Transmission detector (PMT)

2. Inverted microscope in halogen transillumination mode (from top to bottom)

  • light source (halogen lamp)
  • Polarizer (polarizing filter)
  • DIC prism (condenser)
  • Condenser front lens
  • Sample
  • Objective
  • DIC prism at objective
  • Analyzer (polarizing filter)
  • Detector (camera, eye)

3. Upright microscope in halogen transillumination mode (from top to bottom)

  • Detector (camera, eye)
  • Analyzer (polarizing filter)
  • DIC prism at objective
  • Objective
  • Sample
  • Condenser front lens
  • DIC prism (condenser)
  • Polarizer (polarizing filter)
  • light source (halogen lamp)

4. Inverted microscope in laser scanning mode (from top to bottom)

  • Transmission detector (PMT)
  • Analyzer (polarizing filter)
  • DIC prism (condenser)
  • Condenser front lens
  • Sample
  • Objective
  • DIC prism at objective
  • no Polarizer (laser has to provide correctly polarized light)
  • light source (laser)

Prerequisites for DIC observation / recording

1.Koehler alignment of illumination path (transmitted light mode)

  • If objective iris aperture is present: check if open)
  • Focus a sample in transmitted light mode
  • Close field aperture of illumination (first aperture after lamp)
  • Focus field aperture (sharp contrast)
  • Center field aperture
  • Open field aperture (until aperture is no more visible)

2. DIC alignment

  • Remove both DIC prisms from the illumination path
  • Insert polarizer and analyzer into the illumination path
  • Rotate polarizer or analyzer to minimize brightness
  • Insert appropriate objective and condenser DIC prisms
  • Adjust the condenser aperture roughly to the objective's aperture

DIC observation / recording

1. Visual inspection / CCD recording

  • Fine tune objective DIC prism angle (using the dedicated screw or rotation facility at the objective's DIC prism)
  • Switch the slider (tube slider at the eyepiece) to the appropriate beam path setting for CCD recording
  • Adjust camera gain / brightness etc.

2. Laser scanning

  • After DIC alignment (above), remove the polarizer (the laser illumination beam path should provide the correct laser polarization direction at the objective DIC prism. (In case of questions, ask your administrator)
  • Set the beam path to the transmission photomultiplier tube (PMT) recording mode (for details, see the respective microscope's manual)
  • Adjust PMT offset / gain etc.

Note: with some microscopes, there may be problems in case of simultaneous DIC / fluorescence recording. Check the respective microscope's manual). It is important to match the immersion medium between sample and condenser to the immersion medium between lens and sample.

Zur Redakteursansicht