Differential Interference Contrast (DIC)

Prerequisites for DIC observation / recording

1.Koehler alignment of illumination path (transmitted light mode)

  • If objective iris aperture is present: check if open)
  • Focus a sample in transmitted light mode
  • Close field aperture of illumination (first aperture after lamp)
  • Focus field aperture (sharp contrast)
  • Center field aperture
  • Open field aperture (until aperture is no more visible)

2. DIC alignment

  • Remove both DIC prisms from the illumination path
  • Insert polarizer and analyzer into the illumination path
  • Rotate polarizer or analyzer to minimize brightness
  • Insert appropriate objective and condenser DIC prisms
  • Adjust the condenser aperture roughly to the objective's aperture